A Two-Channel Optical Circuit for the Delivery of Incoherent Radiation into the Illumination Channel of a Medical Endoscope


Дәйексөз келтіру

Толық мәтін

Ашық рұқсат Ашық рұқсат
Рұқсат жабық Рұқсат берілді
Рұқсат жабық Тек жазылушылар үшін

Аннотация

An optical circuit for delivering lamp radiation to the light guide of the illumination channel of a medical endoscope was developed. The proposed optical circuit splits the lamp light flux into two parts using systems consisting of a spherical mirror and a lens located opposite each other symmetrically with respect to the lamp. Next, the light fluxes are introduced through focusing cones into intermediate light guides and merged. The resulting light flux is then introduced into the light guide of the illumination channel of a medical endoscope. According to the estimates, the proposed optical system provides delivery of ~25% of xenon-lamp radiation into the light guide of the endoscope’s illumination channel.

Авторлар туралы

I. Kirin

Orenburg State University

Хат алмасуға жауапты Автор.
Email: igkirin@rambler.ru
Ресей, Orenburg

Қосымша файлдар

Қосымша файлдар
Әрекет
1. JATS XML

© Springer Science+Business Media, LLC, part of Springer Nature, 2019