HPLC Determination of Glyphosate, Aminomethylphosphonic Acid, and Glufosinate Using a Hypercarb Porous Graphite Adsorbent


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

A method for the HPLC determination of glyphosate, aminomethylphosphonic acid, and glufosinate using the gradient separation of analytes on a Hypercarb porous graphitized carbon adsorbent and an aqueous solution of ammonium formate and ammonia as a mobile phase is proposed. Analytes are detected using quadrupole and three-quadrupole mass spectrometers. In order to increase the retention of the analytes, the chromatographic column is washed with water before the injection of a sample solution. This procedure results in a three- to fourfold increase in the retention factors of the analytes in comparison with the analogues described in the publications.

About the authors

E. N. Goncharova

Department of Chemistry

Email: mstatkus@gmail.com
Russian Federation, Moscow, 119991

M. A. Statkus

Department of Chemistry

Author for correspondence.
Email: mstatkus@gmail.com
Russian Federation, Moscow, 119991

G. I. Tsizin

Department of Chemistry

Email: mstatkus@gmail.com
Russian Federation, Moscow, 119991

R. N. Selimov

Russian State Center for the Quality and Standardization of Veterinary Medicines and Feeds

Email: mstatkus@gmail.com
Russian Federation, Moscow

Supplementary files

Supplementary Files
Action
1. JATS XML

Copyright (c) 2018 Allerton Press, Inc.