A Prism Mass-Spectrometer for Isotope Analysis of Hydrogen–Helium Mixtures


如何引用文章

全文:

开放存取 开放存取
受限制的访问 ##reader.subscriptionAccessGranted##
受限制的访问 订阅存取

详细

An original scheme of a special magnetic mass-spectrometer for isotope measurements in a mass range of hydrogen–helium mixtures and their main impurities is proposed. The ion optics system contains a two-dimensional magnetic prism supplemented with two identical cylindrical electrostatic capacitors matched so that the system operates in the regime of complete triple focusing of the ion beam in two orthogonal directions and with respect to ion energies. The system ensures high dispersion and allows the mass resolution (no less than 3500) required for analysis at high brightness and a central ion trajectory length on the order of 1 m.

作者简介

L. Gall

Institute for Analytical Instrumentation

编辑信件的主要联系方式.
Email: lngall@yandex.ru
俄罗斯联邦, St. Petersburg, 190103

E. Yakushev

Institute of Nuclear Physics

Email: lngall@yandex.ru
哈萨克斯坦, Almaty, 050032

L. Nazarenko

Institute of Nuclear Physics

Email: lngall@yandex.ru
哈萨克斯坦, Almaty, 050032

A. Antonov

Institute for Analytical Instrumentation; Ioffe Physical Technical Institute

Email: lngall@yandex.ru
俄罗斯联邦, St. Petersburg, 190103; St. Petersburg, 194021

A. Semenov

Advanced Technology R&D Institute of Inorganic Materials

Email: lngall@yandex.ru
俄罗斯联邦, Moscow, 123060

N. Gall

Institute for Analytical Instrumentation; Ioffe Physical Technical Institute

Email: lngall@yandex.ru
俄罗斯联邦, St. Petersburg, 190103; St. Petersburg, 194021

补充文件

附件文件
动作
1. JATS XML

版权所有 © Pleiades Publishing, Ltd., 2018