Mathematical Modeling of Sorption in a Wearable Artificial Kidney with Dialysate Regeneration
- Authors: Bazaev N.A.1, Bizyukov I.O.1, Streltsov E.V.1
-
Affiliations:
- National Research University of Electronic Technology
- Issue: Vol 50, No 5 (2017)
- Pages: 318-320
- Section: Article
- URL: https://journal-vniispk.ru/0006-3398/article/view/234721
- DOI: https://doi.org/10.1007/s10527-017-9646-2
- ID: 234721
Cite item
Abstract
This work is devoted to mathematical modeling of spent dialysate regeneration by sorption. A method for mathematical modeling of sorption of such metabolites as creatinine and uric acid in the sorption columns of a wear able artificial kidney is suggested. The method was tested with different types of activated carbon used in sorption columns. The root-mean-square criterion did not exceed 630 μmol/L.