Abstract
Gaussian profile is conventionally assumed as the probe shape of the incident electron beam in theoretical analysis of dimensional measurements by a scanning electron microscope (SEM). However, it is not suitable for samples with small and tiny structure. In this paper, a model of a focusing electron beam with finite width due to aberration was used in simulating the SEM image of gold particles/balls on a carbon substrate. An effective electron beam shape (EEBS) was displayed and was found that it deviates significantly from the Gaussian profile. The difference between images simulated by Monte Carlo method with ideal electron incident beam, electron beam focusing model and with ideal beam incident then convoluted by Gaussian profile were discussed in detail. Furthermore, the influence of external electric field effect: full extraction and no extraction for imaging were studied.