Determining the Quantum Characteristics of a Photon Counter Using Power-Calibrated Laser Radiation
- Authors: Kovalev A.A.1, Liberman A.A.1, Mikryukov A.S.1, Moskalyuk S.A.1, Ulanovskii M.V.1
-
Affiliations:
- All-Russia Research Institute of Optophysical Measurements (VNIIOFI)
- Issue: Vol 58, No 11 (2016)
- Pages: 1244-1249
- Section: Article
- URL: https://journal-vniispk.ru/0543-1972/article/view/245573
- DOI: https://doi.org/10.1007/s11018-016-0878-4
- ID: 245573
Cite item
Abstract
An experimental setup was devised allowing high-accuracy radiation power measurements in the range of the order of 10−18–10−15 W. The use of a calibrated Fresnel prismatic attenuator in the design minimizes the error in determining the mean number of photos in each radiation pulse. Results are presented for measurements of the positive measure operator for a Hamamatsu H7422-40 photoelectric multiplier.
About the authors
A. A. Kovalev
All-Russia Research Institute of Optophysical Measurements (VNIIOFI)
Author for correspondence.
Email: mrsergik@mail.ru
Russian Federation, Moscow
A. A. Liberman
All-Russia Research Institute of Optophysical Measurements (VNIIOFI)
Email: mrsergik@mail.ru
Russian Federation, Moscow
A. S. Mikryukov
All-Russia Research Institute of Optophysical Measurements (VNIIOFI)
Email: mrsergik@mail.ru
Russian Federation, Moscow
S. A. Moskalyuk
All-Russia Research Institute of Optophysical Measurements (VNIIOFI)
Email: mrsergik@mail.ru
Russian Federation, Moscow
M. V. Ulanovskii
All-Russia Research Institute of Optophysical Measurements (VNIIOFI)
Email: mrsergik@mail.ru
Russian Federation, Moscow
Supplementary files