Calibration of Radiation Sources and Detectors Over a Broad Range of Measurement of the Energy of Optical Radiation
- Authors: Kuvaldin E.V.1
-
Affiliations:
- Vavilov State Optical Institute
- Issue: Vol 61, No 12 (2019)
- Pages: 1181-1186
- Section: Article
- URL: https://journal-vniispk.ru/0543-1972/article/view/246651
- DOI: https://doi.org/10.1007/s11018-019-01567-4
- ID: 246651
Cite item
Abstract
A method of calibration of devices used to measure energy by means of a calorimeter and reference photodiode is described. Experimental results of the calibration are presented. Through the use of the method, energy meters, photodetector devices, and radiation sources may be calibrated over a range more than 108 times the measured radiant energy with the use of time conversion.