A Pulse EPR 25 mT Magnetometer with 10 ppm Resolution
- Autores: Sundramoorthy S.V.1, Epel B.1, Halpern H.J.1
- 
							Afiliações: 
							- Department of Radiation and Cellular Oncology, Center for EPR Imaging in Vivo Physiology, The University of Chicago Medical Center
 
- Edição: Volume 48, Nº 8 (2017)
- Páginas: 805-811
- Seção: Original Paper
- URL: https://journal-vniispk.ru/0937-9347/article/view/247785
- DOI: https://doi.org/10.1007/s00723-017-0902-0
- ID: 247785
Citar
Resumo
A magnetometer designed for permanent magnet manufacturing and operated around 25 mT with 10 ppm absolute accuracy is described. The magnetometer uses pulse electron paramagnetic resonance methodology. The use of a pulsed broadband acquisition allowed reliable measurements in the presence of the magnetic field gradient and in relatively inhomogeneous magnetic fields of un-shimmed magnets.
Sobre autores
Subramanian Sundramoorthy
Department of Radiation and Cellular Oncology, Center for EPR Imaging in Vivo Physiology, The University of Chicago Medical Center
														Email: bepel@uchicago.edu
				                					                																			                												                	Estados Unidos da América, 							5841 S. Maryland Ave, MC1105, Room E-SB-05, Chicago, IL, 60637-1463						
Boris Epel
Department of Radiation and Cellular Oncology, Center for EPR Imaging in Vivo Physiology, The University of Chicago Medical Center
							Autor responsável pela correspondência
							Email: bepel@uchicago.edu
				                	ORCID ID: 0000-0002-1382-7029
				                																			                												                	Estados Unidos da América, 							5841 S. Maryland Ave, MC1105, Room E-SB-05, Chicago, IL, 60637-1463						
Howard Halpern
Department of Radiation and Cellular Oncology, Center for EPR Imaging in Vivo Physiology, The University of Chicago Medical Center
														Email: bepel@uchicago.edu
				                					                																			                												                	Estados Unidos da América, 							5841 S. Maryland Ave, MC1105, Room E-SB-05, Chicago, IL, 60637-1463						