An X-Band Crossed-Loop EPR Resonator


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

A copper X-band (9.22 GHz) cross-loop resonator has been constructed for use with 4 mm sample tubes. The Q for the two resonators is 380 and 350, respectively. The resonator efficiency is about 1 G per square root of watt. Operation has been demonstrated with measurement of T1 by saturation recovery for samples of coal and an immobilized nitroxide.

Sobre autores

George Rinard

School of Engineering and Computer Science and Center for EPR Imaging In Vivo Physiology, University of Denver

Email: geaton@du.edu
Estados Unidos da América, Denver, CO, 80208

Richard Quine

School of Engineering and Computer Science and Center for EPR Imaging In Vivo Physiology, University of Denver

Email: geaton@du.edu
Estados Unidos da América, Denver, CO, 80208

Joseph McPeak

Department of Chemistry and Biochemistry and Center for EPR Imaging In Vivo Physiology, University of Denver

Email: geaton@du.edu
Estados Unidos da América, 2101 E. Wesley Ave., Denver, CO, 80208

Laura Buchanan

Department of Chemistry and Biochemistry and Center for EPR Imaging In Vivo Physiology, University of Denver

Email: geaton@du.edu
Estados Unidos da América, 2101 E. Wesley Ave., Denver, CO, 80208

Sandra Eaton

Department of Chemistry and Biochemistry and Center for EPR Imaging In Vivo Physiology, University of Denver

Email: geaton@du.edu
Estados Unidos da América, 2101 E. Wesley Ave., Denver, CO, 80208

Gareth Eaton

Department of Chemistry and Biochemistry and Center for EPR Imaging In Vivo Physiology, University of Denver

Autor responsável pela correspondência
Email: geaton@du.edu
ORCID ID: 0000-0001-7429-8469
Estados Unidos da América, 2101 E. Wesley Ave., Denver, CO, 80208

Arquivos suplementares

Arquivos suplementares
Ação
1. JATS XML

Declaração de direitos autorais © Springer-Verlag GmbH Austria, 2017