X-ray Absorption Spectroscopy Analysis of Heavy Metals by Means of a Silicon Detector
- Authors: Osadchii S.M.1, Petukhov A.A.1, Dunin V.B.2
-
Affiliations:
- All-Russian Research Institute of Physical-Technical and Radiotechnical Measurements
- Joint Institute for Nuclear Research
- Issue: Vol 13, No 4 (2019)
- Pages: 683-689
- Section: Article
- URL: https://journal-vniispk.ru/1027-4510/article/view/196396
- DOI: https://doi.org/10.1134/S1027451019040116
- ID: 196396
Cite item
Abstract
A Si-PIN detector with longitudinal arrangement of the silicon wafer for detecting X-ray quanta with an energy of up to 100 keV is designed. Its resolution is estimated. An X-ray absorption spectroscopy method for the detection of heavy metals at the K-absorption edges is implemented using the designed detector. The results of measuring the thicknesses of lead and gold foils placed behind a shield by the X-ray absorption spectroscopy method and with a micrometer gauge are compared. The detection limits of heavy metals behind steel or aluminum shields are determined. The measurement error of the metal-foil thicknesses is estimated.
About the authors
S. M. Osadchii
All-Russian Research Institute of Physical-Technical and Radiotechnical Measurements
Author for correspondence.
Email: osm@vniiftri.ru
Russian Federation, Mendeleevo, Moscow oblast, 141570
A. A. Petukhov
All-Russian Research Institute of Physical-Technical and Radiotechnical Measurements
Email: osm@vniiftri.ru
Russian Federation, Mendeleevo, Moscow oblast, 141570
V. B. Dunin
Joint Institute for Nuclear Research
Email: osm@vniiftri.ru
Russian Federation, Dubna, Moscow oblast, 141980
Supplementary files