Vacuum Intensification of Filling Defects with Penetrants under Liquid-Penetrant Testing


Cite item

Full Text

Open Access Open Access
Restricted Access Access granted
Restricted Access Subscription Access

Abstract

The vacuum intensification of the process of filling defects in stainless steel and glass samples with penetrants under liquid-penetrant testing has been studied experimentally. Results indicate a high efficiency of vacuum impregnation with small amounts of penetrants on tested surfaces, especially when detecting flaws with openings of 1–3 μm. It has been established that the sensitivity of testing is increased by 30–100% compared with the traditional technique.

About the authors

N. V. Delenkovskii

Institute of Applied Physics

Author for correspondence.
Email: delenkovsky@iaph.bas-net.by
Belarus, Minsk, 220072

A. B. Gnusin

Institute of Applied Physics

Author for correspondence.
Email: dubel@iaph.bas-net.by
Belarus, Minsk, 220072

Supplementary files

Supplementary Files
Action
1. JATS XML

Copyright (c) 2019 Pleiades Publishing, Ltd.