Calibration of CryojetHT and Cobra Plus Cryosystems used in X-ray diffraction studies
- Authors: Dudka A.P.1, Verin I.A.1, Smirnova E.S.1
-
Affiliations:
- Shubnikov Institute of Crystallography
- Issue: Vol 61, No 4 (2016)
- Pages: 692-696
- Section: Apparatus
- URL: https://journal-vniispk.ru/1063-7745/article/view/190227
- DOI: https://doi.org/10.1134/S1063774516040052
- ID: 190227
Cite item
Abstract
CryoJetHT (Oxford Instruments) and Cobra Plus (Oxford Cryosystems) cryosystems, which are used for sample cooling in X-ray diffraction experiments, have been calibrated. It is shown that the real temperature in the vicinity of the sample differs significantly (the deviation is as high as 8–10 K at low temperatures) from the temperature recorded by authorized sensors of these systems. The calibration results are confirmed by measurements of the unit-cell parameters of GdFe3(BO3)4 single crystal in the temperature range of its phase transition. It is shown that, to determine the real temperature of a sample, one must perform an independent calibration of cryosystems rather than rely on their ratings.