Application of the Monte Carlo method for spectrometer calibration to determine the surface activity of radionuclides deposited on the ground
- Authors: Finkel F.V.1, Rebyakova V.A.1,2, Spirin D.O.2
-
Affiliations:
- Emergency Response Center
- Peter the Great St. Petersburg Polytechnic University
- Issue: Vol 79, No 9-10 (2016)
- Pages: 1424-1428
- Section: Radiation Safety
- URL: https://journal-vniispk.ru/1063-7788/article/view/191157
- DOI: https://doi.org/10.1134/S1063778816090039
- ID: 191157
Cite item
Abstract
The results of efficiency calibration and verification of the in situ method with application of a handheld spectrometer with high-purity germanium detector are presented. The conducted studies show that the calibration of spectrometer efficiency with the use of the Monte Carlo method can be applied for measurement of the surface activity of radionuclides deposited on the ground with an uncertainty of not more than 22%.