Specifics of Preparation of Storage Chambers for Ultracold Neutrons and EDM Spectrometer Sensitivity
- Authors: Lasakov M.S.1, Pirozhkov A.N.1, Serebrov A.P.1
-
Affiliations:
- Petersburg Nuclear Physics Institute, National Research Center Kurchatov Institute
- Issue: Vol 64, No 3 (2019)
- Pages: 436-439
- Section: Experimental Instruments and Technique
- URL: https://journal-vniispk.ru/1063-7842/article/view/203159
- DOI: https://doi.org/10.1134/S1063784219030162
- ID: 203159
Cite item
Abstract
Storage cells for ultracold neutrons (UCNs) are crucial parts of the experimental setup used to measure the electric dipole moment (EDM) of a neutron. The effect of the technology of preparation and sputtering of elements of UCN storage cells on the limiting operating voltage and on the capacity of these cells for long-term operation under high voltage with polarity switching was studied. A voltage magnitude of ±250 kV was obtained in tests of a neutron storage cell for the EDM spectrometer with a cell height of 9 cm.