THE POSSIBILITY TO USE THE ULTRASOUND MONITOR OF NONINVASIVE CONTROL OF HEMODYNAMICS IN NEWBORNS

Cover Page

Cite item

Full Text

Abstract

Functional Doppler echocardiography and estimation of cardiac output using the ultrasound monitor USCOM (Ultrasound Cardiac Output Monitor) are informative and affordable methods of non-invasive control of hemodynamics in newborns.
The study was aimed to determine the significance level of the results of measuring hemodynamics parameters using the non-invasive ultrasound monitor obtained by a beginning operator who lacks the skills of working with ultrasound devices.
The central hemodynamics parameters were determined using the ultrasound monitor of cardiac output non-invasive control USCOM via the left-sided transaortal approach in newborn children staying at the department of resuscitation and intensive therapy. Data were measured by two operators. One of them wasn’t aware of the results obtained by the other investigator. Stroke output, heart rate and cardiac index were chosen to be compared.
Statistical analysis was done using Wilkinson’s coefficient and Spearman’s rank correlation coefficient. Median values were almost the same. In correlation comparison, a high correlation ratio was found for the cardiac rates obtained by operators 1 and 2 (r=0,827; p=0,000; R²=0,68); the stroke volume values were r=0,837; p=0,000; R²=0,7; the values of cardiac output were r=0,781; p=0,000; R²=0,61.
Acquiring practical skills of dealing with USCOM monitor doesn’t need special costs. On-site training under supervision of an instructor is enough.

About the authors

I. V. Boronina

Author for correspondence.
Email: irinaboronina@bk.ru
Заведующая кафедрой анестезиологии и реаниматологии, Воронежский государственный медицинский университет им. Н. Н. Бурденко Минздрава России, 394036, Воронеж, ул.Студенческая, д. 10 Russian Federation

Y. S. Alexandrovich

Email: fake@neicon.ru
Заведующий кафедрой анестезиологии, реаниматологии и неотложной педиатрии факультета послевузовского и дополнительного профессионального образования, Санкт-Петербургский государственный педиатрический медицинский университет Минздрава России, 194110, Санкт-Петербург, ул. Литовская, д. 2 Russian Federation

A. N. Shmakov

Email: fake@neicon.ru
Профессор кафедры анестезиологии и реаниматологии, Новосибирский государственный медицинский университет Минздрава России, 630091, Красный проспект, д. 52 Russian Federation

L. S. Oshanova

Email: fake@neicon.ru
Аспирант кафедры анестезиологии и реаниматологии, Воронежский государственный медицинский университет им. Н. Н. Бурденко Минздрава России, 394036, ул. Студенческая, д. 10 Russian Federation

References

  1. Mertens L., Seri I., Marek J., Arlettaz R., Barkera P., McNamara P., Moon-Grady A. J., Coon P. D., Noory S., Simpson J., Lai W. W. Targeted Neonatal Echocardiography in the Neonatal Intensive Care Unit: Practice Guidelines and Recommendations for Training. European Journal of Echocardiography. 2011; 12: 715–36. PubMed PMID: 21998460.
  2. Lam J. M., Tang C. Emergency physicians can reliably assess patient cardiac output using non-invasive Ultrasonic Cardiac Output Monitor (USCOM). Hon Kong College of Emergence Medicine Scientific Session, 2005. Available from: http://www.uscom.com.au/news/data/HongKongabstract.pdf.
  3. Dey I., Sprivuls P. Emergency physicians can reliably assess emergency department patient cardiac output using the USCOM continuous wave Doppler cardiac output monitor. Emergency Medicine Australasia. 2005; 17: 193–9. PubMed PMID: 15953218.
  4. Stewart G. M., Nguyen H. B., Kim T. Y., Jauregui J., Hayes S. R., Corbett S. Inter-Rater Reliability for Noninvasive Measurement of Cardiac Function in Children. Pediatric Emergency Care. 2008; 24 (7): 433–7. PubMed PMID: 18580818.
  5. Kager С. C. M., Dekker G. A., Stam M. C. Measurement of cardiac output in normal pregnancy by a non-invasive twodimensional independent Doppler device. Australian and New Zealand Journal of Obstetrics and Gynecology. 2009; 49: 142–4. PubMed PMID: 19441163.
  6. Patel N., Dodsworth M., Mills J. F. Cardiac output measurement in newborn infants using the ultrasonic cardiac output monitor: an assessment of agreement with conventional echocardiography, repeatability and new user experience. Arch Dis Child Fetal Neonatal Ed. 2010; 96 (3): 206–11. PubMed PMID: 20605971.
  7. Meyer S., Todd D. A. Assessment of portable continuous wave Doppler ultrasound (ultrasonic cardiac output monitor) for cardiac output measurements in neonates. Journal of Pediatrics and Child Health. 2009; 45 (7-8): 464–8. PubMed PMID: 19712182.

Supplementary files

Supplementary Files
Action
1. JATS XML

Согласие на обработку персональных данных с помощью сервиса «Яндекс.Метрика»

1. Я (далее – «Пользователь» или «Субъект персональных данных»), осуществляя использование сайта https://journals.rcsi.science/ (далее – «Сайт»), подтверждая свою полную дееспособность даю согласие на обработку персональных данных с использованием средств автоматизации Оператору - федеральному государственному бюджетному учреждению «Российский центр научной информации» (РЦНИ), далее – «Оператор», расположенному по адресу: 119991, г. Москва, Ленинский просп., д.32А, со следующими условиями.

2. Категории обрабатываемых данных: файлы «cookies» (куки-файлы). Файлы «cookie» – это небольшой текстовый файл, который веб-сервер может хранить в браузере Пользователя. Данные файлы веб-сервер загружает на устройство Пользователя при посещении им Сайта. При каждом следующем посещении Пользователем Сайта «cookie» файлы отправляются на Сайт Оператора. Данные файлы позволяют Сайту распознавать устройство Пользователя. Содержимое такого файла может как относиться, так и не относиться к персональным данным, в зависимости от того, содержит ли такой файл персональные данные или содержит обезличенные технические данные.

3. Цель обработки персональных данных: анализ пользовательской активности с помощью сервиса «Яндекс.Метрика».

4. Категории субъектов персональных данных: все Пользователи Сайта, которые дали согласие на обработку файлов «cookie».

5. Способы обработки: сбор, запись, систематизация, накопление, хранение, уточнение (обновление, изменение), извлечение, использование, передача (доступ, предоставление), блокирование, удаление, уничтожение персональных данных.

6. Срок обработки и хранения: до получения от Субъекта персональных данных требования о прекращении обработки/отзыва согласия.

7. Способ отзыва: заявление об отзыве в письменном виде путём его направления на адрес электронной почты Оператора: info@rcsi.science или путем письменного обращения по юридическому адресу: 119991, г. Москва, Ленинский просп., д.32А

8. Субъект персональных данных вправе запретить своему оборудованию прием этих данных или ограничить прием этих данных. При отказе от получения таких данных или при ограничении приема данных некоторые функции Сайта могут работать некорректно. Субъект персональных данных обязуется сам настроить свое оборудование таким способом, чтобы оно обеспечивало адекватный его желаниям режим работы и уровень защиты данных файлов «cookie», Оператор не предоставляет технологических и правовых консультаций на темы подобного характера.

9. Порядок уничтожения персональных данных при достижении цели их обработки или при наступлении иных законных оснований определяется Оператором в соответствии с законодательством Российской Федерации.

10. Я согласен/согласна квалифицировать в качестве своей простой электронной подписи под настоящим Согласием и под Политикой обработки персональных данных выполнение мною следующего действия на сайте: https://journals.rcsi.science/ нажатие мною на интерфейсе с текстом: «Сайт использует сервис «Яндекс.Метрика» (который использует файлы «cookie») на элемент с текстом «Принять и продолжить».