The Optimal Surgical Needle for Tendon Suture: Cutting Edge or Reverse Cutting Edge?

封面

如何引用文章

全文:

详细

Background. Achieving  a  durable  connection  between  the  lacerated  tendon  ends  is  difficult.  The  outcome  of  treatment depends on many factors. Several authors consider the properties of the surgical needle used for suturing the tendon to be important. The aim of the study— to compare the strength of the tendon suture applied with the conventional cutting edge and reverse cutting edge surgical needles in the experiment.

Materials and Methods.We used porcine tendons for the experiment. The tendon fragments were divided into 2 groups of 20 tendons each. On all 40 tendons, the same type of “injury” of the tendon was simulated — using a scalpel. In the first group, the interrupted suture of the tendon was applied with a cutting edge surgical needle, in the second group — reverse cutting edge. Laboratory tests of the tendon sutures strength were performed on the improvised stand.

Results.In the first (suture made with a cutting needle edge), diastasis of 2 mm was determined at an average load of 1219.5 g (m = ±76.56, where «m» is the representativeness error). Complete suture failure occurred at an average load of 1770.8 g (m = ±100.02). In this group, the thread rupture was not recorded. In the second group (a suture made with a reverse cutting edge needle), diastasis occurs with an average load of 1754.75 g (m = ±77.32). Complete suture failure occurred at an average load of 2571.25 (at m= ± 103.78). In three cases, the thread ruptured. In the second group (reverse cutting edge needle), the tendon suture strength was statistically significantly higher than in the first group.

 Conclusion. The tendon suture strength depends on the surgical needle properties. In tendons  reconstruction  the  reverse  cutting  edge  needle  use  is  more  preferable  compared  to  the  conventional  cutting  edge needle use.

作者简介

A. Zolotov

Far Eastern Federal University

Email: dalexpk@gmail.com
ORCID iD: 0000-0002-0045-9319

Aleksandr  S.  Zolotov —  Dr.  Sci.  (Med.),  Professor

Vladivostok

俄罗斯联邦

S. Isokov

Far Eastern Federal University

编辑信件的主要联系方式.
Email: sultanisakov5@gmail.com
ORCID iD: 0000-0001-7971-3174

Sultonali  Isokov 

Vladivostok

俄罗斯联邦

A. Isokova

Far Eastern Federal University

Email: aziza19-99@mail.ru
ORCID iD: 0000-0001-8344-6769

Aziza Isokova

Vladivostok

俄罗斯联邦

参考

  1. Jong J.P. de, Nguyen J.T., Sonnema A.J.M., Nguyen E.C., Amadio P.C., Moran S.L. The Incidence of Acute Traumatic Tendon Injuries in the Hand and Wrist: A 10-year Populationbased Study. Clin Orthop Surg. 2014;6(2):196-202. doi: 10.4055/cios.2014.6.2.196.
  2. Золотов А.С., Зеленин В.Н., Сороковиков В.А. Хирургическое лечение повреждений сухожилий сгибателей пальцев кисти. Иркутск: РИО НЦ РВХ ВСНЦ СО РАМН, 2006. 110 с.
  3. Byrne M., Aly A. The Surgical Needle. Aesthet Surg J. 2019;39(Suppl_2):S73-S77. doi: 10.1093/asj/sjz035.4. Lam W.L., Garrido A, Vandermeulen J., Fagan M.J., Stanley P.R.W. Cutting or round-bodied needles for tendon repair. J Hand Surg Br. 2003;28(5):475-477. doi: 10.1016/s0266-7681(03)00169-4.
  4. Construction Flexor Tendon Repair Assessment Tool. Creation of Reproducible Testing Method. ASSH. Available from: https://www.assh.org/s/surgical-simulation.
  5. Tang J.B. New Developments Are Improving Flexor Tendon Repair. Plast Reconstr Surg. 2018;141(6):1427-1437. doi: 10.1097/PRS.0000000000004416.
  6. Iwanaga y., Morizaki y., Uehara K., Tanaka S., Sakai T., Saito T. Robust Suture Combination for Rat Flexor Tendon Repair Model. J Hand Surg Global Online.2020:354-358. doi: 10.1016/j.jhsg.2020.08.004.
  7. Winters S.C., Gelberman R.H., Woo S.L., Chan S.S., Grewal R., Seiler J.G. 3rd. The effects of multiplestrand suture methods on the strength and excursion of repaired intrasynovial flexor tendons: a biomechanical study in dogs. J Hand Surg Am. 1998;23(1):97-104. doi: 10.1016/s0363-5023(98)80096-8.
  8. Wong y.R., Tay S.C. A Biomechanical Study of a Novel Asymmetric 6-Strand Flexor Tendon Repair Using Porcine Tendons. Hand (N Y). 2018;13(1):50-55. doi: 10.1177/1558944716685829.
  9. Kang G.H., Wong y.R., Lim R.Q., Loke A.M., Tay S.C. Cyclic Testing of the 6-Strand Tang and Modified Lim-Tsai Flexor Tendon Repair Techniques. J Hand Surg Am. 2018;43(3):285.e1-285.e6. doi: 10.1016/j.jhsa.2017.08.014.
  10. Грицюк А.А., Середа А.П. Ахиллово сухожилие. М.: РАЕН, 2010. 313 с.
  11. Wallace S.J., Mioton L.M., Havey R.M., Muriuki M.G., Ko J.H. Biomechanical Properties of a Novel Mesh Suture in a Cadaveric Flexor Tendon Repair Model. J Hand Surg Am. 2019;44(3):208-215. doi: 10.1016/j.jhsa.2018.11.016.
  12. Galvez M.G., Comer G.C., Chattopadhyay A., Long C., Behn A.W., Chang J. Gliding Resistance After Epitendinous-First Repair of Flexor Digitorum Profundus in Zone II. J Hand Surg Am.2017;42(8):662.e1-662.e9. doi: 10.1016/j.jhsa.2017.04.013.
  13. Lim R.Q.R., Wong y.-R., Loke A.M.K., Tay S.-C. A cyclic testing comparison of two flexor tendon repairs: asymmetric and modified Lim-Tsai techniques. J Hand Surg Eur. 2018;43(5):494-498. doi: 10.1177/1753193418758828.
  14. Tang y.Q., Mao W.F., Wu y.F. A new and accurate method to quantify gap formation in tendon repairs. J Hand Surg Eur. 2013;38(7):808-809. doi: 10.1177/1753193412455286.

补充文件

附件文件
动作
1. JATS XML

版权所有 © Traumatology and Orthopedics of Russia, 2023

Согласие на обработку персональных данных с помощью сервиса «Яндекс.Метрика»

1. Я (далее – «Пользователь» или «Субъект персональных данных»), осуществляя использование сайта https://journals.rcsi.science/ (далее – «Сайт»), подтверждая свою полную дееспособность даю согласие на обработку персональных данных с использованием средств автоматизации Оператору - федеральному государственному бюджетному учреждению «Российский центр научной информации» (РЦНИ), далее – «Оператор», расположенному по адресу: 119991, г. Москва, Ленинский просп., д.32А, со следующими условиями.

2. Категории обрабатываемых данных: файлы «cookies» (куки-файлы). Файлы «cookie» – это небольшой текстовый файл, который веб-сервер может хранить в браузере Пользователя. Данные файлы веб-сервер загружает на устройство Пользователя при посещении им Сайта. При каждом следующем посещении Пользователем Сайта «cookie» файлы отправляются на Сайт Оператора. Данные файлы позволяют Сайту распознавать устройство Пользователя. Содержимое такого файла может как относиться, так и не относиться к персональным данным, в зависимости от того, содержит ли такой файл персональные данные или содержит обезличенные технические данные.

3. Цель обработки персональных данных: анализ пользовательской активности с помощью сервиса «Яндекс.Метрика».

4. Категории субъектов персональных данных: все Пользователи Сайта, которые дали согласие на обработку файлов «cookie».

5. Способы обработки: сбор, запись, систематизация, накопление, хранение, уточнение (обновление, изменение), извлечение, использование, передача (доступ, предоставление), блокирование, удаление, уничтожение персональных данных.

6. Срок обработки и хранения: до получения от Субъекта персональных данных требования о прекращении обработки/отзыва согласия.

7. Способ отзыва: заявление об отзыве в письменном виде путём его направления на адрес электронной почты Оператора: info@rcsi.science или путем письменного обращения по юридическому адресу: 119991, г. Москва, Ленинский просп., д.32А

8. Субъект персональных данных вправе запретить своему оборудованию прием этих данных или ограничить прием этих данных. При отказе от получения таких данных или при ограничении приема данных некоторые функции Сайта могут работать некорректно. Субъект персональных данных обязуется сам настроить свое оборудование таким способом, чтобы оно обеспечивало адекватный его желаниям режим работы и уровень защиты данных файлов «cookie», Оператор не предоставляет технологических и правовых консультаций на темы подобного характера.

9. Порядок уничтожения персональных данных при достижении цели их обработки или при наступлении иных законных оснований определяется Оператором в соответствии с законодательством Российской Федерации.

10. Я согласен/согласна квалифицировать в качестве своей простой электронной подписи под настоящим Согласием и под Политикой обработки персональных данных выполнение мною следующего действия на сайте: https://journals.rcsi.science/ нажатие мною на интерфейсе с текстом: «Сайт использует сервис «Яндекс.Метрика» (который использует файлы «cookie») на элемент с текстом «Принять и продолжить».