Polyelectrolyte-based enzymatic diagnosticum with precipitation detection system for urea assay in solution


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

A polyelectrolyte-based enzymatic diagnosticum with a precipitation detection system that can be used as a biosensor was created. The detection method was based on the change in polyelectrolyte microcapsule weight with respect to the urea content. The possibility of biosensor reutilization was demonstrated. The appropriate ionic precipitator causing precipitation of insoluble carbonate within the microcapsules and the optimal microcapsule titre were found. In the solution of monovalent anions (chlorides), the activity of encapsulated urease was shown to increase monotonically as the square root of the ionic strength depending on the elevation of the salt content. The activity drastically increased in a narrow concentration interval (0.6–0.8 mM) of divalent anions (sulfates) and reached the level of the native enzyme activity.

Sobre autores

A. Dubrovsky

Institute of Theoretical and Experimental Biophysics

Email: tikhonenkosa@gmail.com
Rússia, Pushchino, Moscow oblast, 142290

L. Shabarchina

Institute of Theoretical and Experimental Biophysics

Email: tikhonenkosa@gmail.com
Rússia, Pushchino, Moscow oblast, 142290

S. Tikhonenko

Institute of Theoretical and Experimental Biophysics

Autor responsável pela correspondência
Email: tikhonenkosa@gmail.com
Rússia, Pushchino, Moscow oblast, 142290

Arquivos suplementares

Arquivos suplementares
Ação
1. JATS XML

Declaração de direitos autorais © Pleiades Publishing, Inc., 2017