Selection of Parameters of Bandpass Filtering of the ECG Signal for Heart Rhythm Monitoring Systems
- Authors: Fedotov A.A.1
-
Affiliations:
- Samara State Aerospace University
- Issue: Vol 50, No 2 (2016)
- Pages: 114-118
- Section: Article
- URL: https://journal-vniispk.ru/0006-3398/article/view/234461
- DOI: https://doi.org/10.1007/s10527-016-9600-8
- ID: 234461
Cite item
Abstract
This work is devoted to the bandpass frequency filtering of the ECG signal in heart rhythm monitoring systems against the background of disturbances and noises of various origins. It was suggested to optimize the procedure for selection of the bandwidth and the type of bandpass filter using the criteria for minimization of uncertainty in the measurement of the R-R intervals. Two different approaches to the bandpass filtering were considered: use of a second-order analog active filter and of an eighth-order digital filter.
About the authors
A. A. Fedotov
Samara State Aerospace University
Author for correspondence.
Email: fedoaleks@yandex.ru
Russian Federation, Samara
Supplementary files