The Effect of Transmitter Coil Size on the Optimal Implantation Depth of Receiver Coil in Transcutaneous Inductive Energy Transfer Systems
- Авторлар: Danilov A.A.1, Mindubaev E.A.1, Selishchev S.V.1
-
Мекемелер:
- National Research University of Electronic Technology
- Шығарылым: Том 51, № 5 (2018)
- Беттер: 354-357
- Бөлім: Article
- URL: https://journal-vniispk.ru/0006-3398/article/view/235278
- DOI: https://doi.org/10.1007/s10527-018-9747-6
- ID: 235278
Дәйексөз келтіру
Аннотация
The effect of patient’s individual characteristics on the output of wireless transcutaneous inductive energy transfer systems has been studied. Numerical modeling shows that small changes in the implantation depth due to patients’ individual characteristics (thickness of skin, subcutaneous fat layer, and bones) can lead to significant changes in output power and efficiency of the energy transfer. Our study shows that the effect of these changes can be compensated by optimizing the size of the external (transmitter) coil.
Авторлар туралы
A. Danilov
National Research University of Electronic Technology
Email: edmindubaev@gmail.com
Ресей, Zelenograd, Moscow
E. Mindubaev
National Research University of Electronic Technology
Хат алмасуға жауапты Автор.
Email: edmindubaev@gmail.com
Ресей, Zelenograd, Moscow
S. Selishchev
National Research University of Electronic Technology
Email: edmindubaev@gmail.com
Ресей, Zelenograd, Moscow