Key Comparison of the Standards for the Volt of the Russian Federation and the Republic of Belarus Using a Quantum Comparison Standard
- Authors: Katkov A.S.1, Chernyaev P.A.2
-
Affiliations:
- Mendeleev All-Russia Research Institute of Metrology (VNIIM)
- Belarus State Institute of Metrology (BelGIM)
- Issue: Vol 59, No 4 (2016)
- Pages: 444-448
- Section: International Collaboration
- URL: https://journal-vniispk.ru/0543-1972/article/view/245882
- DOI: https://doi.org/10.1007/s11018-016-0987-0
- ID: 245882
Cite item
Abstract
The procedure and results of key comparisons of the national voltage standards of the Russian Federation and the Republic of Belarus using a 10 V quantum comparison standard are discussed. The comparisons agree with the reference value obtained from the BIPM voltage standard. The relative difference was 0.9·10–10 with a combined relative standard uncertainty of 2.6·10–10.
About the authors
A. S. Katkov
Mendeleev All-Russia Research Institute of Metrology (VNIIM)
Author for correspondence.
Email: a.s.katkov@vniim.ru
Russian Federation, St. Petersburg
P. A. Chernyaev
Belarus State Institute of Metrology (BelGIM)
Email: a.s.katkov@vniim.ru
Belarus, Minsk
Supplementary files