On locating a vacuum leak with the indicator fiber particle method


Citar

Texto integral

Acesso aberto Acesso aberto
Acesso é fechado Acesso está concedido
Acesso é fechado Somente assinantes

Resumo

A mathematical model is presented for the interaction of a fiber indicator particle with a free molecular gas flow. An inverse problem for locating a leak is formulated and solved using particular trajectories and kinematic characteristics of the indicator’s motion. Diagnostic signs that are sufficient for locating the leak are revealed.