Disengagement of Polyethylene Insert Locking Mechanism in Modular Tibial Components for Knee Arthroplasty: A Case Report
- Authors: Chugaev C.V.1, Kuliaba T.A.1, Petukhov A.I.1, Martynenko A.I.1
-
Affiliations:
- Vreden National Medical Research Center of Traumatology and Orthopedics
- Issue: Vol 30, No 1 (2024)
- Pages: 120-128
- Section: Case Reports
- URL: https://journal-vniispk.ru/2311-2905/article/view/255322
- DOI: https://doi.org/10.17816/2311-2905-17411
- ID: 255322
Cite item
Abstract
Background. Modular tibial components for knee arthroplasty are used in the majority of modern knee replacement systems. Despite a number of limitations, there are many aspects that make these types of implants indispensable for orthopedic surgeons.
Aim — to demonstrate possible risks associated with a modular polyethylene liner with metal locking clip using a clinical case as an example.
Case description. We present a case of primary total knee arthroplasty in a 70-year-old female patient. The surgery was performed by an experienced surgical team and resulted in good early radiologic and functional treatment outcome. After discharge, approximately 10 days after surgery, the patient developed knee pain. Control X-rays showed migration of the metal pin locking the polyethylene insert. The patient underwent an emergency revision surgery with replacement of the clip. The authors analyze possible causes of this complication and ways of its prevention.
Conclusion. Migration of the insert locking element and dislocation of the insert in locked systems are quite rare complications of the knee arthroplasty. Their causes are soft tissue imbalance of the knee joint during arthroplasty and a number of technical errors. The very fact of using modular components of the joint is a predisposing factor for the disassociation of these modules.
Full Text
##article.viewOnOriginalSite##About the authors
Chugaev V. Chugaev
Vreden National Medical Research Center of Traumatology and Orthopedics
Author for correspondence.
Email: dr.chugaev@gmail.com
ORCID iD: 0000-0001-5127-5088
Cand. Sci. (Med.)
Russian Federation, St. PetersburgTaras A. Kuliaba
Vreden National Medical Research Center of Traumatology and Orthopedics
Email: taraskuliaba@mail.ru
ORCID iD: 0000-0003-3175-4756
Dr. Sci. (Med.)
Russian Federation, St. PetersburgAleksey I. Petukhov
Vreden National Medical Research Center of Traumatology and Orthopedics
Email: drpetukhov@yandex.ru
ORCID iD: 0000-0002-2403-6521
Cand. Sci. (Med.)
Russian Federation, St. PetersburgAnastasiia I. Martynenko
Vreden National Medical Research Center of Traumatology and Orthopedics
Email: martynenko.anst@yandex.ru
ORCID iD: 0009-0005-0525-6456
Russian Federation, St. Petersburg
References
- Castellarin G., Bori E., Menon A., Innocenti B. The effect of different insert design congruencies on the kinematics of a mobile bearing TKA: A cadaveric study. J Orthop. 2022;34:89-93. doi: 10.1016/j.jor.2022.07.018.
- Stulberg S.D., Goyal N. Which Tibial Tray Design Achieves Maximum Coverage and Ideal Rotation: Anatomic, Symmetric, or Asymmetric? An MRI-based study. J Arthroplasty. 2015;30(10):1839-1841. doi: 10.1016/j.arth.2015.04.033.
- Longo U.G., Ciuffreda M., D’Andrea V., Mannering N., Locher J., Denaro V. All-polyethylene versus metal-backed tibial component in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2017;25(11):3620-3636. doi: 10.1007/s00167-016-4168-0.
- AbuMoussa S., White C.C. 4th, Eichinger J.K., Friedman R.J. All-Polyethylene versus Metal-Backed Tibial Components in Total Knee Arthroplasty. J Knee Surg. 2019;32(8):714-718. doi: 10.1055/s-0039-1683979.
- Łapaj Ł, Mróz A., Kokoszka P., Markuszewski J., Wendland J., Helak-Łapaj C., et al. Peripheral snap-fit locking mechanisms and smooth surface finish of tibial trays reduce backside wear in fixed-bearing total knee arthroplasty. Acta Orthop. 2017;88(1): 62-69. doi: 10.1080/17453674.2016.1248202.
- Conditt M.A., Ismaily S.K., Alexander J.W., Noble P.C. Backside wear of modular ultra-high molecular weight polyethylenetibial inserts. J Bone Joint Surg Am. 2004;86(5):1031-1037. doi: 10.2106/00004623-200405000-00022.
- Thienpont E. Failure of tibial polyethylene insert locking mechanism in posterior stabilized arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2013;21(12):2685-2688. doi: 10.1007/s00167-012-2018-2.
- Chen C.E., Juhn R.J., Ko J.Y. Dissociation of polyethylene insert from the tibial baseplate following revision total knee arthroplasty. J Arthroplasty. 2011;26(2):339.e11-339.e13. doi: 10.1016/j.arth.2010.04.016.
- Sisko Z.W., Teeter M.G., Lanting B.A., Howard J.L., McCalden R.W., Naudie D.D., et al. Current Total Knee Designs: Does Baseplate Roughness or Locking Mechanism Design Affect Polyethylene Backside Wear? Clin Orthop Relat Res. 2017;475(12):2970-2980. doi: 10.1007/s11999-017-5494-3.
- Hepinstall M.S., Rodriguez J.A. Polyethylene subluxation: a radiographic sign of locking mechanism failure after modular total knee arthroplasty. J Arthroplasty. 2011;26(1):98-102. doi: 10.1016/j.arth.2009.10.020.
- Sanders A.P., Raeymaekers B. The effect of polyethylene creep on tibial insert locking screw loosening and back-out in prosthetic knee joints. J Mech Behav Biomed Mater. 2014;38:1-5. doi: 10.1016/j.jmbbm.2014.06.002.
- Чугаев Д.В., Кравцов Е.Д., Корнилов Н.Н., Куляба Т.А. Анатомо-биомеханические особенности латерального отдела коленного сустава и связанные с с ними технические аспекты одномыщелкового эндопротезирования: лекция. Травматология и ортопедия России. 2023;29(2):144-158. doi: 10.17816/2311-2905-2042. Chugaev D.V., Kravtsov E.D., Kornilov N.N., Kulyaba T.A. Anatomical and Biomechanical Features of the Lateral Compartment of the Knee and Associated Technical Aspects of Unicompartmental Knee Arthroplasty: Lecture. Traumatology and Orthopedics of Russia. 2023;29(2):144-158. doi: 10.17816/2311-2905-2042. (In Russian).
- Rapuri V.R., Clarke H.D., Spangehl M.J., Beauchamp C.P. Five cases of failure of the tibial polyethylene insert locking mechanism in one design of constrained knee arthroplasty. J Arthroplasty. 2011;26(6):976.e21-976.e24. doi: 10.1016/j.arth.2010.07.013.
- Cho W.S., Youm Y.S. Migration of polyethylene fixation screw after total knee arthroplasty. J Arthroplasty. 2009;24(5):825.e5-825.e9. doi: 10.1016/j.arth.2008.07.011.
- Scott R.D., Chmell M.J. Balancing the posterior cruciate ligament during cruciate-retaining fixed and mobile-bearing total knee arthroplasty: description of the pull-out lift-off and slide-back tests. J Arthroplasty. 2008;23(4):605-608. doi: 10.1016/j.arth.2007.11.018.
- Migon E.Z., de Freitas G.L, Rodrigues M.W., de Oliveira G.K., de Almeida L.G., Schwartsmann C.R. Spontaneous dislocation of the polyethylene component following knee revision arthroplasty: case report. Rev Bras Ortop. 2014;50(1):114-116. doi: 10.1016/j.rboe.2014.12.002.
- Jindal S., Bansal V., Ahmed M. Disengagement of tibial insert locking pin in total knee arthroplasty - A rare failure case report. J Clin Orthop Trauma. 2022;33:101996. doi: 10.1016/j.jcot.2022.101996.
- Nachtnebl L., Tomáš T., Apostolopoulos V., Pazourek L., Mahdal M. Long-Term Results of Total Knee Replacement Using P.F.C. Sigma System with an All-Polyethylene Tibial Component. Acta Chir Orthop Traumatol Cech. 2021;88(6):412-417. doi: 10.55095/achot2021/061. (In Czech).
- Brihault J., Navacchia A., Pianigiani S., Labey L., De Corte R., Pascale V., et al. All-polyethylene tibial components generate higher stress and micromotions than metal-backed tibial components in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2016;24(8):2550-2559. doi: 10.1007/s00167-015-3630-8.
- Wasielewski R.C., Parks N., Williams I., Surprenant H., Collier J.P., Engh G. Tibial insert undersurface as a contributing source of polyethylene wear debris. Clin Orthop Relat Res. 1997;(345):53-59.
- Wasielewski R.C. The causes of insert backside wear in total knee arthroplasty. Clin Orthop Relat Res. 2002;(404): 232-246. doi: 10.1097/00003086-200211000-00037.
- Norgren B., Dalén T., Nilsson K.G. All-poly tibial component better than metal-backed: a randomized RSA study. Knee. 2004;11(3):189-196. doi: 10.1016/S0968-0160(03)00071-1.
- Kumar V., Hasan O., Umer M., Baloch N. Cemented all-poly tibia in resource constrained country, affordable and cost-effective care. Is it applicable at this era? Review article. Ann Med Surg (Lond). 2019;47:36-40. doi: 10.1016/j.amsu.2019.09.010.
- Apostolopoulos V., Nachtnebl L., Mahdal M., Pazourek L., Boháč P., Janíček P., et al. Clinical outcomes and survival comparison between NexGen all-poly and its metal-backed equivalent in total knee arthroplasty. Int Orthop. 2023;47(9):2207-2213. doi: 10.1007/s00264-023-05772-3.
- Gudnason A., Hailer N.P., W-Dahl A., Sundberg M., Robertsson O. All-Polyethylene Versus Metal-Backed Tibial Components-An Analysis of 27,733 Cruciate-Retaining Total Knee Replacements from the Swedish Knee Arthroplasty Register. J Bone Joint Surg Am. 2014;96(12):994-999. doi: 10.2106/JBJS.M.00373.
- Apostolopoulos V., Tomáš T., Boháč P., Marcián P., Mahdal M., Valoušek T., et al. Biomechanical analysis of all-polyethylene total knee arthroplasty on periprosthetic tibia using the finite element method. Comput Methods Programs Biomed. 2022;220:106834. doi: 10.1016/j.cmpb.2022.106834.
- Lewis P.L., Robertsson O., Graves S.E., Paxton E.W., Prentice H.A., W-Dahl A. Variation and trends in reasons for knee replacement revision: a multi-registry study of revision burden. Acta Orthop. 2021;92(2):182-188. doi: 10.1080/17453674.2020.1853340.
- Тихилов Р.М., Корнилов Н.Н., Куляба Т.А., Филь А.С., Дроздова П.В. Принципы создания и функционирования регистров артропластики коленного сустава. Вестник военно-медицинской академии. 2014;1(45):220-226. Tikhilov R.M., Kornilov N.N., Kulyaba T.A., Fil A.S., Drozdova P.V. Principles of creation and functioning of knee arthroplasty register. Bulletin of the Russian Military Medical Academy. 2014;1(45):220-226. (In Russian).
- Jensen C.L., Petersen M.M., Jensen K.E., Therbo M., Schrøder H.M. Outcome of isolated tibial polyethylene insert exchange after uncemented total knee arthroplasty: 27 patients followed for 8-71 months. Acta Orthop. 2006;77(6):917-920. doi: 10.1080/17453670610013222.
Supplementary files