Mathematical simulation of a membrane capacitive compensation pressure transducer
- 作者: Eikhval’d A.I.1, Sadkovskaya I.V.2, Eikhval’d T.A.2
-
隶属关系:
- St. Petersburg State University
- D.I. Mendeleyev Institute for Metrology
- 期: 卷 42, 编号 8 (2016)
- 页面: 769-771
- 栏目: Article
- URL: https://journal-vniispk.ru/1063-7850/article/view/200343
- DOI: https://doi.org/10.1134/S1063785016080071
- ID: 200343
如何引用文章
详细
The operation of a membrane capacitive pressure transducer with electrostatic compensation of the membrane deformation has been mathematically simulated. It is shown that the theory describing these instruments is incorrect and the error in measuring pressure by them may be as high as 25–29% of the measured value, which makes them inappropriate for use in pressure metrology.